12 month warranty Request a Quote Technical Support
Sign In

Product Specifications

Product Description

The AccuMac AM1762 Secondary SPRT features an accuracy of ±0.006°C at 0.01°C, a short term stability of ±0.002°C and a very low drift rate of less than 19mK after 500 hours. The sensing element is designed to protect the platinum sensing wire from contamination at high temperatures, giving the device a high level of stability and repeatability in performance. A uniquely designed support structure and filling material provides an excellent balance between the hysteresis effect, mechanical shock, and thermal shock performance.

AccuMac
Compare

Secondary SPRT - Wide Range | AccuMac AM1762

-200 to 670°C, ±0.006°C accuracy, ITS-90
MPN/Series: AM1762

Product Specifications

Brand
AccuMac
Model Number
AM1762
Type
Secondary SPRT
Temperature Range
-200 to 670°C
Accuracy
±0.006°C at 0.01°C
Drift
<19mK after 500 hours
Documents
Product Manual, Product Specs
Secondary SPRT - Wide Range | AccuMac AM1762