12 month warranty Request a Quote Technical Support
Sign In

Product Specifications

Product Description

The TSEM puts a magnetic float switch and a temperature sensor on one flange. The vessel gains level limits and a temperature reading through one process connection, so there is one gasket, one cable entry and one leak test instead of two.

The float switches the level limits, and up to two limit value inputs are available, so one instrument covers high and low alarm duties, while the temperature output is a potential-free relay contact that can be wired into the same panel as the level contacts. Up to two limit value inputs can be set as high and low alarms for one tank.

The process window is -50 to 125 °C. Pressure stops at 1 bar. The sliding tube comes in stainless steel or brass, chosen by the aggressiveness of the medium.

It is not a transmitter. It reports limit states and a temperature contact, so a level trend or a 4-20 mA signal needs a different instrument.

The 1 bar limit is the constraint to check first, because a pressurised vessel is outside this design. Above 125 °C the unit is outside its range as well.

Engler Steuer-Mess- und Regeltechnik
Compare

Engler TSEM Magnetic Float Level Switch, Flange Mount with Temperature Sensor

-50 to 125 C, 0 to 1 bar, two limits
MPN/Series: TSEM

Product Specifications

Brand
Engler Steuer-Mess- und Regeltechnik
Part Number
TSEM
Type
Magnetic float level switch with integrated temperature sensor
Media
Liquids
Mounting
Flange connection
Process Temperature
-50 to +125 °C
Process Pressure
0 to 1 bar
Limit Values
Up to 2 limit value inputs
Temperature Output
Potential-free relay contact
Sliding Tube
Stainless steel or brass, optional
Engler TSEM Magnetic Float Level Switch, Flange Mount with Temperature Sensor