Product Specifications
Manufacturer: Rosemount
Product Number: Model 404 Endurance Series
Product Type: Low Flow Contacting Conductivity Sensor with Integrated Flow Cell
Available Cell Constants: 0.01, 0.1, 1.0, and 10/cm
Electrode Material (0.01/0.1/1.0/cm): Concentric titanium electrodes
Electrode Material (10/cm): Graphite electrodes
Insulator Material: PEEK
Body Material: PVC or stainless steel (stainless steel version is disassemblable for cleaning)
Seal Material: EPDM O-rings
Temperature Compensation: Integrated platinum RTD
Process Connection: 3/8 inch barbed tubing connector
Recommended Conductivity Range: Up to 20,000 µS/cm (clean, non-corrosive liquids)
Core Features: Small holdup volume, rapid response to conductivity/temperature changes
Primary Applications: High Purity Water, Clean Cooling Water, Ultrapure Water Systems
Documents: Product Manual , Product Specs
Product Description
The Rosemount Model 404 Endurance Low Flow Conductivity Sensors combine the conductivity sensor with a flow cell. The design has a small holdup volume, so response to sudden changes in process conductivity and temperature is rapid. Model 404 sensors are available with a PVC or stainless steel body. The stainless steel version can be disassembled for cleaning. The PVC version cannot be taken apart.
ENDURANCE sensors are contacting conductivity sensors. They are available in cell constants of 0.01, 0.1, 1.0, and 10/cm. The choice of cell constant depends on conductivity. High conductivity samples require larger cell constants. Consult the analyzer Product Data Sheet for recommended ranges and accuracy.
Sensors with 0.01, 0.1 and 1.0/cm cell constants have concentric titanium electrodes separated by a PEEK insulator. EPDM O-rings seal the internal parts of the sensor from the process liquid. Sensors with 10/cm cell constants have graphite electrodes and an epoxy body.
Electrolytic conductivity is a strong function of temperature, and conductivity readings are typically converted to the value at a reference temperature. To measure temperature, ENDURANCE sensors use a platinum RTD. In metal electrode sensors the RTD is inside the inner electrode. In the epoxy body sensor, the RTD is in a titanium thermowell protruding from the end of the sensor.
The Model 404 sensors are intended for the determination of electrolytic conductivity in applications ranging from high purity water to clean cooling water. The sensors are ideal for use in clean, non-corrosive liquid having conductivity less than about 20,000 µS/cm. For dirty or corrosive samples or for samples having high conductivity, a toroidal sensor such as the Model 228 or 226 is recommended.
Features
- 0.01 and 0.1/cm cell constants
- 3/8 inch barbed tubing connector
-
Wetted materials:
- Electrodes; titanium
- Insulator; PEEK
- O-rings; EPDM
- PVC or stainless steel body
Low Flow Conductivity Sensor - Ultra-Fast Response, High Purity Water | Rosemount 404 Endurance
Product Specifications
Manufacturer: Rosemount
Product Number: Model 404 Endurance Series
Product Type: Low Flow Contacting Conductivity Sensor with Integrated Flow Cell
Available Cell Constants: 0.01, 0.1, 1.0, and 10/cm
Electrode Material (0.01/0.1/1.0/cm): Concentric titanium electrodes
Electrode Material (10/cm): Graphite electrodes
Insulator Material: PEEK
Body Material: PVC or stainless steel (stainless steel version is disassemblable for cleaning)
Seal Material: EPDM O-rings
Temperature Compensation: Integrated platinum RTD
Process Connection: 3/8 inch barbed tubing connector
Recommended Conductivity Range: Up to 20,000 µS/cm (clean, non-corrosive liquids)
Core Features: Small holdup volume, rapid response to conductivity/temperature changes
Primary Applications: High Purity Water, Clean Cooling Water, Ultrapure Water Systems
Documents: Product Manual , Product Specs