Product Specifications
- Manufacturer
- Edgetech Instruments
- Product Type
- Portable Dew Point Monitor
- Part Number
- 1500
- Sensor
- Two-Stage or Three-Stage Chilled Mirror
- Dew / Frost Point Range
- -75 to 75°C
- Accuracy
- ±0.20°C (±0.36°F)
- Sample Connection
- 1/4in Compression Fitting
- Sample Flow Rate
- 0.5 to 5 SCFH; Integral Flow Meter
- Analog Output
- 4-20mA or 0-5VDC (scalable for ranges -50 to 50°C or -75 to 75°C)
- Power
- AC model or with rechargeable batteries and charger
- Features
- Self-contained system, no field assembly required, integral flow meter, protective carrying case
- Documents
- Product Specs
Product Description
The Edgetech 1500 Portable Dew Point Monitor is a fully self-contained system that brings laboratory-grade chilled mirror accuracy to the field. Its integrated design eliminates the need to carry and assemble separate components like control boxes, sensors, cables, and pumps, making it exceptionally easy to deploy for spot checks or troubleshooting.
Operation is straightforward: simply connect the sample line, adjust the flow using the integral flow meter, and power on. Within minutes, the 1500 begins delivering accurate, drift-free dew point data over a wide measurement range of -75 to 75°C . Its rugged construction is built to withstand the demands of field use, yet it maintains the high precision required for critical measurements.
The monitor's fundamental chilled mirror measurement principle ensures results that are more accurate (±0.20°C) than those calculated from separate moisture and temperature sensors. NIST traceability is standard. The 1500 is available in both AC-powered and rechargeable battery versions, and a protective case is included for easy and safe transportation .
Resources & Documentation
Portable Dew Point Monitor - Self-Contained, Chilled Mirror | Edgetech 1500
Product Specifications
- Manufacturer
- Edgetech Instruments
- Product Type
- Portable Dew Point Monitor
- Part Number
- 1500
- Sensor
- Two-Stage or Three-Stage Chilled Mirror
- Dew / Frost Point Range
- -75 to 75°C
- Accuracy
- ±0.20°C (±0.36°F)
- Sample Connection
- 1/4in Compression Fitting
- Sample Flow Rate
- 0.5 to 5 SCFH; Integral Flow Meter
- Analog Output
- 4-20mA or 0-5VDC (scalable for ranges -50 to 50°C or -75 to 75°C)
- Power
- AC model or with rechargeable batteries and charger
- Features
- Self-contained system, no field assembly required, integral flow meter, protective carrying case
- Documents
- Product Specs