12 month warranty Request a Quote Technical Support
Sign In

Product Specifications

Product Description

The Endress+Hauser Promass A 100 is a single-tube Coriolis flow meter, DN 1 to DN 4, for mass flow measurement at ultra-low rates.

The measuring principle does not depend on viscosity or density, so the reading is mass flow without a medium-specific conversion. The measuring tube is a single straight tube, so it drains under gravity when the line is emptied.

The tube is 1.4539 (904L) or Alloy C22, and the connections are 1.4404 (316/316L) or Alloy C22. Process pressure reaches 400 bar. The medium may run to 205 °C.

Mass flow error is 0.1% of reading for liquids and 0.5% for gases, volume flow is 0.1%, density is 0.0005 g/cm³, and the measuring range is 0 to 450 kg/h, which is where this meter differs from a DN 8 Promass: the small tube is the point, not a reduced range.

The transmitter housing is ultra-compact and rated IP69K, so it can be washed down. The meter needs no inlet or outlet run. A local display is available, and the integrated web server gives access without extra software. Heartbeat Technology provides verification in place.

The limits are the tube and the process. A DN 1 tube passes a narrow stream, so solids or fibres that would pass a DN 25 meter will block this meter, and the 400 bar figure is an upper limit that depends on the process connection. The meter measures mass directly, so it does not need a density input for the flow reading.

Endress+Hauser
Compare

Endress+Hauser Proline Promass A 100 Coriolis Flow Meter, DN 1-4, 400 bar

DN 1-4, 400 bar, ±0.1%
MPN/Series: 8P1B

Product Specifications

Manufacturer
Endress+Hauser
Product Number
Proline Promass A 100 / 8A1B
Type
Ultra-low flow Coriolis flow meter
Nominal Diameter
DN 1 to 4 (1/24 to 1/8 inch)
Mass Flow Error (Liquid)
±0.1%
Volume Flow Error (Liquid)
±0.1%
Mass Flow Error (Gas)
±0.5%
Density Error (Liquid)
±0.0005 g/cm³
Measuring Range
0 to 450 kg/h (0 to 16.5 lb/min)
Medium Temperature Range
-50 to +205 °C (-58 to +401 °F)
Max Process Pressure
400 bar (5800 psi)
Wetted Materials
Measuring tube 1.4539 (904L) or Alloy C22; connections 1.4404 (316/316L) or Alloy C22
Transmitter Housing
Ultra-compact; optional stainless steel hygienic housing
Ingress Protection
IP69K
Local Display
Available
Endress+Hauser Proline Promass A 100 Coriolis Flow Meter, DN 1-4, 400 bar