12 month warranty Request a Quote Technical Support
Sign In

Product Specifications

Product Description

The E+H Memosens CLS16E Digital Conductivity Sensor features a certified hygienic design conforming to FDA and USP Class VI. Memosens CLS16E measures conductivity in pure and ultrapure water applications, and is certified with quality certificate and EN 10204 3.1, with additional compliance from EHEDG. Extended storage of calibration and process allows for more accurate trend identification and expanded predictive maintenance, with secure digital transfer of historical data via Memosens 2.0 technology. The non-contact, inductive signal transmission method of the CPS16E ensures maximum process safety. The hygienic process connections available with the CLS16E allow for installation in pipes or flow vessels. The sensor is serializable and autoclavable, resists cleaning in place (CIP) and sterilization in place (SIP).

E+H Memosens CLS16E Digital Conductivity Sensor applications:

  • Monitoring of ion exchangers
  • Reverse osmosis
  • Distillation
  • Electrodeionization
  • WFI (water for injection) in the pharmaceutical industry

Features

  • Measuring principle:
    • Conductive
    • 2 electropolished stainless steel electrodes
  • Measurement range: k=0.1: 0.04 to 500 µS/cm
  • Process pressure:
    • 13 bar at 20°C (188 psi at 68°F)
    • 9 bar at 120°C (130 psi at 248°F)
  • Process temperature:
    • -5 to 120°C (23 to 248°F)
    • For Sterilization: max. 150°C at 5 bar (Max. 302°F at 73 psi)
  • Process connection(s):
    • ISO2852 Clamp 1-1/2in / 2in
    • Neumo BioControl D50
    • Varivent N DN50-125
  • Temperature sensor: Pt1000
  • Material(s):
    • Sensor depending on order execution:
      • Electropolished, stainless steel 1.4435 (AISI 316L)
      • PEEK
    • Sealing depending on order execution:
      • Form seal ISOLAST (FFKM)
      • Form seal EPDM
  • IP68
  • ATEX, NEPSI, CSA, IECEx, INMETRO, and EAC Ex approvals
  • Additional certifications:
    • Material certification 3.1
    • EHEDG certified, hygienic design
    • CoC ASME BPE
    • Conformity to cGMP derived
Endress+Hauser
Compare

Pharmaceutical Hygienic Conductivity Sensor - WFI Monitoring, cGMP Compliant | E+H Memosens CLS16E

EHEDG certified, non-contact signal transmission
MPN/Series: Memosens CLS16E

Product Specifications

Manufacturer
Endress+Hauser (E+H)
Product Number
Memosens CLS16E
Product Type
Digital Hygienic Conductivity Sensor
Measuring Principle
Conductive, 2 electropolished stainless steel electrodes
Core Measuring Range
k=0.1: 0.04 to 500 µS/cm
Process Temperature
-5 to 120°C (23 to 248°F); max. 150°C at 5 bar for sterilization (max 302°F at 73 psi)
Process Pressure
13 bar at 20°C (188 psi at 68°F); 9 bar at 120°C (130 psi at 248°F)
Process Connections
ISO2852 Clamp 1-1/2in / 2in, Neumo BioControl D50, Varivent N DN50-125
Temperature Sensor
Pt1000
Wetted Materials
Electropolished 316L stainless steel (1.4435), PEEK; Seals: ISOLAST (FFKM), EPDM
Ingress Protection
IP68
Core Technology
Memosens 2.0 digital technology (calibration/process data storage, non-contact inductive signal transmission)
Hazardous Area Approvals
ATEX, NEPSI, CSA, IECEx, INMETRO, EAC Ex
Hygienic Certifications
Material certification 3.1, EHEDG certified hygienic design, CoC ASME BPE, cGMP derived conformity, FDA, USP Class VI
Core Features
Sterilizable/autoclavable, CIP/SIP compatible
Primary Applications
Ion Exchanger Monitoring, Reverse Osmosis, Distillation, Electrodeionization, WFI (Water for Injection) in Pharmaceutical Industry
Documents
Product Manual , Product Specs
Pharmaceutical Hygienic Conductivity Sensor - WFI Monitoring, cGMP Compliant | E+H Memosens CLS16E