The Michell Instruments DG Series Humidity Generators provide flexible, cost-effective solutions for calibration applications based on the volumetric mixing of dry and wet gases. Available in two models, the instruments can be controlled either manually or automatically to suit a wide range of requirements.

DG-2 Humidity Generator: Features two stages of gas-flow mixing, enabling generation of dew points down to -103°F (-75°C). Its key strengths are ease of use and flexibility in manually generating accurate target dew points by fine-tuning the gas mix via flow metering valves. Infinite mixing is achievable within its working range, providing exceptional versatility for low-humidity calibration applications.

DG-3 Humidity Generator: The entry-level, single-stage mixing generator operated by manually mixing two gas streams using flow control valves. It provides flows up to 5 Nl/min (10.6 scfh) and generates dew points ranging from -40 to 68°F (-40 to 20°C), making it suitable for routine calibration requirements.

Both models offer high stability and repeatability with fast response to changes in set-point. They serve as complete dew-point calibration systems for laboratory environments where manual control and precision flow measurement are sufficient for application needs.

Michell Instruments

Humidity Generators - Manual Mixing, -75 to 20°C | Michell DG Series

Manual humidity generators
MPN/Series: DG-2 / DG-3

Product Specifications

Manufacturer:Michell Instruments

Product Type:Humidity Generator

Part Number:DG-2 / DG-3

Principle:Volumetric mixing of dry and wet gases

DG-2 Humidity Range:-75 to 20°C (-103 to 68°F)

DG-3 Humidity Range:-40 to 20°C (-40 to 68°F)

Mixing Stages:DG-2: 2 stages; DG-3: 1 stage

Control:Manual selection via flow control valves

Flow (DG-3):Up to 5 Nl/min (10.6 scfh)

Operation:Precision flowmeter or mass flow controller operation

Features:High stability and repeatability, fast response to set-point changes, infinite mixing achievable within working range (DG-2)

Documents: Product Manual, Product Specs